TEMPERATURE STANDARDS:
TYPE S/AS /ES
PRECISION, ULTRASTABLE NTC STANDARDS
DESCRIPTION:
Thermometrics temperature standards consist of
ultra-stable thermistor probes assembled into thin
wall stainless steel housings with shielded extension
leads. The thermistors used receive special process-
ing to ensure long term stability. All thermistor tem-
perature standards are ruggedly constructed and are
suitable for liquid immersion.
APPLICATIONS:
Thermometrics Thermistor standards are rugged,
precision sensors suitable for use as secondary or
working temperature standards for all laboratory
metrology applications. They generally are not af-
fected by shock and vibration and, consequently,
are also suitable for field use. Thermometrics Tem-
perature standards fill the need for low cost temper-
ature standards for general laboratory and hospital
use, clinical applications and process temperature
measurements. Special versions are available for
military and space use. Standards are also available
for other temperature ranges in a variety of sizes
and enclosures. Our Applications Engineering Staff
can assist you with your specific requirements.
DATA:
CONFIGURATION: Temperature standards are en-
closed in thin-wall stainless tubes, welded closed at
one end. The shielded cable is encapsulated into
the tube to provide a solid moisture-proof seal. Full
immersion of the metal portion of the standard is
permissible. The calibration data is given for an
immersion depth of 8" on 9" housings, and 4" on
41/2" long housings.
TEMPERATURE RANGE: The Type “S” and Type
“AS” are designed for operation over the range of
0°C to 60°C. The Type “ES” is rated for 0°C to
100°C.
STABILITY: The stability of each temperature stan-
dard is guaranteed for a period of one year. When
properly used, the following stability ratings apply:
Type “AS” 0.002° C/year
Type “S” 0.005° C/year
Type “ES” 0.005° C/year
CALIBRATION: Precision calibration, traceable to
the National Institute of Standards and Technology,
is provided for all temperature standards. A com-
puter generated table in increments of 0.01°C is
furnished with each calibration based on the inter-
polation formula, RT=exp (A0+ A1/T + A2/T2+
A3/T3). The constants for the formula are obtained
from a polynomial regression performed on the cal-
ibration data obtained. Over the range of 0°C to
60°C, calibration is performed at the triple point of
water (0.01°C) and 15°C, 25°C, 30°C, 37°C, 50°C,
and 60°C. For the range of 0°C to 100°C, calibra-
tions are performed at the triple point of water,
25°C, 30°C, 37°C, 60°C, 80°C, and 100°C. Two-
wire calibrations are performed using a wheatstone
Bridge calibrated to an accuracy of better than
0.005%. Four-wire calibrations are based on a
comparison technique using a ratio bridge having
an accuracy of 0.0002%. All resistance measure-
ments are referenced to standard resistors calibrat-
ed by NIST. All temperature measurements are
made using a standard platinum resistance ther-
mometer which has been calibrated by NIST.
RESISTANCE VS. TEMPERATURE CHARACTERISTIC: The nominal resistance values are shown below:
TYPE RESISTANCE IN OHMS
O°C 25°C 60°C 100°C
“AS” & “S” 14250 5000 1458 –
“AS” & “S” 11400 4000 1166 –
“ES” 28500 10000 2915 925